Wavelength accuracy
The closeness of an instrument's indicated wavelength position to a traceable reference wavelength.
Sample absorption changes with wavelength. Even with stable photometric readings, measuring at the wrong spectral position biases peak interpretation and quantification at specified wavelengths.
Establishing a wavelength scale from known positions
Wavelength accuracy describes how closely the indicated position of a spectral line or absorption band agrees with a traceable reference. Verification commonly uses emission lines of known position or certified absorption references, determining deviations at specified bandwidth, scan settings and environmental conditions.
NIST's spectrophotometric reference systems support traceability for transmittance, absorbance and wavelength scales separately, and emphasize using reference materials within specified conditions. A certified band position may be defined for a particular spectral bandwidth and peak-position algorithm, so the calculation method cannot be changed arbitrarily.
How a wavelength offset affects absorbance
Near the top of a broad, flat absorption band, a small wavelength offset may have little effect on absorbance. Near a steep spectral edge or narrow peak, the same offset produces a larger photometric difference. The effect depends on the spectral slope and cannot be judged from wavelength error alone.
Nucleic acid and protein experiments usually read specified wavelengths and use neighboring wavelengths to assess purity or background. If the entire wavelength scale shifts, multiple readings are taken at the wrong positions and ratio interpretation may also be affected.
Separate accuracy from repeatability
A reference peak appearing at the same position in ten consecutive measurements indicates good short-term repeatability, but may still be consistently displaced from its certified position. Conversely, an average close to the reference with large variations between measurements is not stable.
Grating or monochromator positioning, temperature changes, calibration models, spectral bandwidth, sampling interval, instrument line shape and peak-position algorithms all affect the result. Passing one wavelength point is insufficient to establish accuracy over the full band; choose multiple reference features covering the range used by the method.
Wavelength accuracy ensures that photometry is read at the correct position; photometric accuracy ensures that the value at that position is close to the reference. Together they support interpretable spectral results.