Why are purity ratios often unstable at low concentration?

Purity ratios divide one absorbance by another. At low concentrations, sample signals may approach blank variability; a small denominator makes the ratio particularly sensitive to small changes.

Inspect individual absorbances and replicate blanks before interpreting the ratio. Repeated cleanup solely to improve a ratio may lose more sample and weaken the signal further.

To quantify dilute nucleic acids, consider a fluorescence assay suited to the target and concentration range, calibrated as instructed. Fluorescence concentration does not replace integrity checks or assessment of suitability for downstream reactions.

Request a NANO X1 trial

Trials are currently available only in mainland China.

Main measurement needs